T1 / R1 relaxation analysis

 

Main Menu:  Modules->T1 relaxation analysis

This module generates output images where the pixel intensities represents longitudinal relaxation times or rates (T1/R1). The input images can be based on one of the three following sequences:

  • Inversion recovery (IR) with multiple (two or more) inversion times (TIs).
  • Saturation recovery with two or more delay times (TDs)
  • Spoiled gradient echo (SPGR) with two or more flip angles (FAs)


Inversion (IR) - and saturation  recovery (SR) sequences
:

IR/SR are sequences containing a 90 (SR) - or 180 deg (IR)  pre-pulse followed by images acquired at multiple delay  (TD) or inversion times (TI) either as a single shot or as multiple acquisitions: The relaxation values are then obtained by performing a non-linear least squares fit to the following expression:

SI = A(1 -K exp(-t*R1))  

for 'real' images (i.e. when the pixel SI can take on both positive and negative values) or: 

SI = A | (1 -K exp(-t* R1)) | + C

for modulus images (when negative signal intensities are converted to its modulus value). In both cases, A and C are constants, K is a variable (equal to 2 for ideal inversion recovery and 1 for ideal saturation recovery), t is the delay- (TD) or  inversion time (TD/TI) and R1 is the longitudinal relaxation rate (=1/T1).  The constant C accounts for the presence of a constant offset in the relaxation curve due to the presence of noise so that the SI never reaches zero (only applicable for modulus IR images)

Multiple flip angle  Spoiled Gradient Echo (SPGR) sequence:

For this sequence type  T1 or R1 are estimated based on spoiled gradient echo signal expression:

SI =A sin(a)(1-exp(-TR/T1)/(1-cos(a)exp(-TR/T1)

where A is a constant, a is the flip angle (FA) and  TR is the repetition time
Two or more flip angles are needed and the resulting quality of the relaxation maps depends strongly on the actual difference in SI for the different FAs used.  It is therefore important to choose FAs which gives a large range of SIs for the T1-values of interest.

T1 Relaxation analysis settings

Sequence specifications

Here you specify which sequence (imaging method) is used to generate the data from which T1 relaxation times  will be calculated . Three sequence types can be specified in the Sequence type drop-down menu:

  • Inversion Recovery (IR)
  • Saturation recovery (SR)
  • Multiple flip angle SPGR

If  sequence type is set to either SR or IR then multiple sequence related options can be set in the IR sequence option panel:

  • IR data type: This can either be Modulus or Real. Se above for details.
  • Apply Look-Locker correction: Corrects the calculated T1/R1 values for the effect of signal saturation in Look-Locker type sequences (multiple RF excitation pulses after a single inversion pulse). The flip angle must be specified when this option is enabled and the option is only valid for IR curves.
  • Correct for imperfect inversion pulse: When T1 is estimated from an Inversion Recovery (IR) sequence, the formula used is: SI(TI) = A[1- B exp(-TI/T1)] where TI is the inversion time, A is a constant reflecting equilibrium magnetization and B is a value reflecting the magnitude of the inversion (180 deg) pulse. For a 'perfect' IR sequence B=2, but the value may deviate from 2 when the B1-excitation field is imperfect, causing the value of B to vary across the field of view. When this option is enabled, the parameter B is allowed to vary (i.e. included as a parameter to be fitted). This will account for variations in the inversion pulse, but also increases the number of degrees of freedom in the fitting procedure, which can result in greater variation in the resulting R1/T1 estimation. Note that when the option 'Apply Look-Locker correction' is enabled, correction for imperfect inversion pulse is always applied, since this is part of the Look-Locker correction algorithm.
  • Baseline offset: When enabled, the constant C in the signal expression above is made a model parameter to account for the fact that the signal level never reaches zero for modulus IR curves (due to image noise). This option is not used for SR images or for real IR data.

Inversion recovery and saturation recovery sequence parameters:

  • Constant TI/TD spacing: Selected if a constant inversion or delay time spacing is used. Min TI/TD  is the time of the first inversion/saturation time and Delta TI/TD  is the spacing of successive time-points  after the first TI/TD  time. All times are in msec.
  • Edit TI/TD values: The inversion- or delay times are specified in a list. This would be used in situations where the TI/TD values  are sampled at uneven intervals (and not specified in the DICOM header).
  • Ignore Dicom values: If TI values are present in the DICOM header, but one wishes to use other values.

Multiple flip-angle SPGR sequence parameters:

  • Constant flip angle increment/decrement: Selected if a constant flip angle increment or decrement is used. First FA is the flip angle of the first image and FA increment/decrement is the constant change in flip angle for each consecutive image in the series. All values are in degrees.
  • Edit flip angles: The flip angles used are specified in a list. This would be used in situations where a non-constant flip angle increment or decrement is used (and the values are not specified in the DICOM header).
  • Repetition time: Specifies the repetition time (TR) of the SPGR sequence (see sequence expression above).
  • Ignore Dicom values: If valid flip angles and TR values are present in the DICOM header, but one wishes to use other values.

Pre-process and Analysis

Noise reduction options:

  • Apply spatial smoothing:  Apply a nearest neighbour smoothing kernel to the raw data prior to curve fitting.
  • Apply temporal smoothing:  The dynamic time signal is smooted (low-pass filtered), to reduce effects of noise and spikes in the dynamic signal response. This smoothing does not affect spatial resolution but may reduce the ability to detect rapid signal changes. If temporal smoothing is selected the degree of smoothing can be varied using the slider.
  • Eliminate spikes: Remove abnormally high (spiky) output values.

Noise Thresholding:

  • Auto Detect: Automatically determine the noise level in the input images (default). The noise level can be modified using the slider.
  • Show noise level cutoff: Show the selected noise threshold as red pixels in the input images. All pixels colored red are excluded from the analysis.
  • Eliminate non-connected voxels (brain extraction): Removes areas that are not connected to the rest of the brain.

Analysis:

It is possible to use different weigthing for the different datapoints. Variable weighting can be useful in order to apply less weight during the curve fitting procedure to data points known to be affected by noise - or to ensure good curve fit for specific parts of the curve. For instance, in an IR sequence, the data-points close to zero magnetization values may be more prone to noise and could be given less weighting. Or alternatively, very high amplitude data points can be given reduced weighting to obtain better fits for data close to zero values.

Output

Here you specify options related to how the parametric images are displayed.

Output Image types: 

  • R1 relaxation rate map: pixel values represent inverse relaxation times (1/T1) in units of 1/sec.
  • T1 relaxation rate map: pixels values represent relaxation times in units of msec. 
  • M0 map: This map represents the relative equilibrium magnetization obtained from the curve fit. In the case of IR and SR sequences, this is obtained from the intensity that is estimated for zero inversion- or saturation time. For multi-flip angle sequences, the value of Mo represents an arbitrary scaling factor needed to fit the equation for spoiled gradient echo sequences to the measured signal variation.
  • Chi-square map: An additional output image is generated which represents the pixel-wise  'goodness of fit' (described by the chi-square parameter resulting from the least squares fit) 

Related topics:

T2 / R2 relaxation analysis

NordicImagingLab logo